Welcome to Shanghai! Our main products: magnetic shutter level gauge, ultrasonic and magnetic level gauge
86-021-56901458
Address:No. 107 Nanshan Road, Shanghai, Jingan District
Tel:86-021-56901458
Fax:86-021-33250243
Email:13817864195@139.com

The integration of explosion-proof heat resistance

Usually the integration of explosion-proof resistance and display instrument, recording instrument, …

Description:

A. Product introduction

Usually the integration of explosion-proof resistance and display instrument, recording instrument, such as the electronic computer, 4 ~ 20 ma output.Direct measurement within the scope of the 0 ~ 1800 ℃ in the process of producing all kinds of liquid, steam and gas medium, as well as solid surface temperature.

2. Product main features

Two wire output 4 ~ 20 ma, strong anti-interference ability;
Save the compensation conductor and temperature transmitter installation cost;
Measuring range is large;
The cold end temperature automatic compensation, nonlinear correction circuit.

3. Working principle

Integration of explosion-proof heat resistance by using the theory of flame-proof clearance, when an explosion cavity, can remove from heat and cooling by joint clearance, make after the explosion flame temperature less than outside the cavity, thereby for temperature measurement.Thermocouple (resistance) production of thermoelectric potential (resistance) after temperature transmitter of bridge produce unbalance signal, after amplification into 4 ~ 20 ma dc signal to the instrument, the instrument will show the corresponding temperature value.

4. Technical parameters

Execution standard
IEC584 IEC1515 IEC751 JB/T5518-91 JB/T7391-94

5. The temperature measurement range and tolerance (thermocouple)

model
Degree,
Allowed error and accidentally material level
Class I
Class II
Tolerance values
℃ temperature range
Tolerance values
℃ temperature range
WRNB
K
Plus or minus 1.5 ℃
- 40 ~ + 375
Plus or minus 2.5 ℃
- 40 ~ + 333
±0.004│t│
375 ~ 1000
±0.0075│t│
333 ~ 1200
WRMB
N
Plus or minus 1.5 ℃
- 40 ~ + 375
Plus or minus 2.5 ℃
- 40 ~ + 333
±0.004│t│
375 ~ 1000
±0.0075│t│
333 ~ 1200
WREB
E
Plus or minus 1.5 ℃
- 40 ~ + 375
Plus or minus 1.5 ℃
- 40 ~ + 333
±0.004│t│
375 ~ 800
±0.004│t│
333 ~ 900
WRFB
J
Plus or minus 1.5 ℃
- 40 ~ + 375
Plus or minus 1.5 ℃
- 40 ~ + 333
±0.004│t│
375 ~ 750
±0.004│t│
333 ~ 750
WRCB
T
Plus or minus 1.5 ℃
- 40 - ~ + 125
+ / - 1 ℃
- 40 ~ + 333
±0.004│t│
125 ~ 350
±0.0075│t│
133 ~ 350
WRPB
S
+ / - 1 ℃
0 ~ + 1100
Plus or minus 2.5 ℃
0 ~ 600
±[1+0.003(t-1100)]
1100 ~ 1600
±0.0025│t│
600 ~ 1600


Note: thermal components measured temperature t absolute value.

Temperature measurement range and tolerance (heat resistance)


model
Degree,
℃ temperature range
Precision grade
Allowable deviation
WZPB
Pt100
- 200 ~ + 500
A grade
±(0.15+0.002|t|)
Class B
±(0.30+0.005|t|)
WZCB
Cu50
- 50 ~ + 100
-
±(0.30+0.006|t|)
Cu100


Note: thermal components measured temperature t absolute value.

Explosion-proof level:

Flame-proof type: dIIBT4, dIICT5 dIICT6
Intrinsically safe type: iaIICT6

6. Other parameters

Output signals: 4 ~ 20 ma, 250 Ω load resistance, 100 Ω transmission conductor resistance
Output method: two wire
Tolerance level: 0.1;0.2;0.5
Power supply: 24 v DC + / - 10%
Protection grade: IP65

Thermal response time

When the temperature appear step change, the instrument of current output signal changes to correspond to the step change of 50% of the time required, usually expressed as a tau 0.5 when temperature transmitter does not exceed the stable step response time of thermocouple (resistance) of stable thermal response time tau 0.5 five points, with a thermocouple (resistance) thermal response time as the instrument of thermal response time;
When the temperature transmitter does not exceed the stable step response time of thermocouple (resistance) 0.5 half of the stable thermal response time tau, the temperature transmitter thermal response time is used as the instrument of the thermal response time.